Chapter 6
Troubleshooting
6.9 LAN Communications
No. Symptom
Check
Remedy
1
The error message
"VS-PC host not
connect" appears.
Check if the
icon appears in the
lower right of the screen.
Check the setting in [Storage Setting].
Set to On to automatically send
data to a PC, otherwise set to Off.
2
The error message
"ECG-PC host not
connect" appears.
Check if the
icon appears in the
lower right of the screen.
Check the setting in [Storage Setting].
Set to On to automatically send
data to a PC, otherwise set to Off.
Check if the
icon appears in the
lower right of the screen.
Check [Order Setting] window in initial
setting/maintenance.
Set operation to [Online].
Enter the IP address and port
number of the order PC.
Check the network settings for
VS-1500 and conduct the [ping Test]
in the maintenance menu.
1. An error occurs immediately.
2. The error occurs again.
1. Correctly enter the IP address of
the order PC.
2. Do not connect the VS-1500
directly to the hospital LAN or
other network. If Ethernet hubs
will be used, use only hubs that
we recommend.
Check if the LAN cable is loose or
broken.
Connect the cable correctly.
Replace the LAN cable if broken.
The hub has not been turned on.
Turn it on.
Check if D14 on the main board lights.
(
☞
10.1.1.C3)
The LAN cable is not properly
connected. Connect the cable
correctly.
3
The error message
"Data cannot send to
Order-PC host"
appears.
Check if D15 and D16 lights during
PC connection.
(
☞
10.1.1.C3)
Check the network settings for
VS-1500 and conduct the [ping
test] in the maintenance menu.
Check the network settings for
VS-1500 and conduct the [ping Test]
in the maintenance menu.
1. An error occurs immediately.
2. The error occurs again.
1. Correctly enter the IP address
and port number of the order
PC.
2. Do not connect the VS-1500
directly to the hospital LAN or
other network. If Ethernet hubs
will be used, use only hubs that
we recommend.
Check if the LAN cable is loose or
broken.
Connect the cable correctly.
Replace the LAN cable if broken.
Check the power supply of the hub
and turn on the power.
Check if D14 on the main board lights.
(
☞
10.1.1.C3)
The LAN cable is not properly
connected. Connect the cable
correctly.
4
The error message
"Data cannot send to
VS-PC host" appears.
The error message
"Data cannot send to
ECG-PC host"
appears.
Check if D15 and D16 lights during
PC connection.
(
☞
10.1.1.C3)
Check the network settings for
VS-1500 and conduct the [ping
test] in the maintenance menu.
5
The error message
"File send error"
appears.
Check if the LAN cable is loose or
broken.
Connect the cable correctly.
Replace the LAN cable if broken.
6
The error message "A
communication error is
occurred" appears.
Check [Network and Master ID
Setting] window in initial
setting/maintenance.
Set Master ID request to [Host].
6-14
Confidential
Copying Prohibited
Summary of Contents for VaSera VS-1500
Page 12: ...Confidential Copying Prohibited x ...
Page 18: ...0 Chapter 1 Features 1 6 Confidential Copying Prohibited ...
Page 32: ...Chapter 4 Names of Parts of Equipment 4 6 Confidential Copying Prohibited ...
Page 42: ...Chapter 5 Maintenance 5 10 Confidential Copying Prohibited Diagonal Line Stripe ...
Page 178: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 3 ...
Page 180: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 5 ...
Page 191: ...Chapter 10 Part Location Diagram and Checkpoints 10 16 Confidential Copying Prohibited ...
Page 207: ...Chapter 11 Spare Parts List 11 16 Confidential Copying Prohibited ...