Chapter
10
Part
Location Diagram and Checkpoints
10.3 LCDTP IF BOARD
10.3.1 Part Location Diagram
10-14
Confidential
Copying Prohibited
10.3.2 Power Supply and Signal Checks
1.
+
3.3V
Allowable voltage is +3.3 ±0.165V.
Check voltage at C12.
2. +5V
Allowable voltage is +5.0 ±0.25V.
Check voltage at C1.
3. +12V
Allowable voltage is +12 ±0.6V.
Check at pin 1 on CN4.
4.
Clock
This is the LCD panel clock signal (25.175 MHz).
Check at R26 or NF23.
Component side, part location diagram
R26
、
NF23
Solder side, part location diagram
+
3.3V
C1
C2
+
5.0V
+
12V
Summary of Contents for VaSera VS-1500
Page 12: ...Confidential Copying Prohibited x ...
Page 18: ...0 Chapter 1 Features 1 6 Confidential Copying Prohibited ...
Page 32: ...Chapter 4 Names of Parts of Equipment 4 6 Confidential Copying Prohibited ...
Page 42: ...Chapter 5 Maintenance 5 10 Confidential Copying Prohibited Diagonal Line Stripe ...
Page 178: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 3 ...
Page 180: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 5 ...
Page 191: ...Chapter 10 Part Location Diagram and Checkpoints 10 16 Confidential Copying Prohibited ...
Page 207: ...Chapter 11 Spare Parts List 11 16 Confidential Copying Prohibited ...