Chapter 5
Maintenance
Confidential
Copying Prohibited
5-29
5.4.2 Checking Cuff Pressure
Use this function to check the pressure sensor used in cuff measurements.
Press the
(Menu) key to open the [Top Menu], then touch [Initial Setting/Maintenance]
Æ
[Maintenance]
Æ
[OK (
↵
)]
Æ
[Next]
Æ
[
Manufacturer Maintenance
]
Æ
[3]
Æ
[3]
Æ
[9]
Æ
[4]
Æ
[
↵
]
Æ
[Pump Unit] to open the [Pump Unit ] window.
CAUTION
The [Pump Unit] window is a factory maintenance tool. Do not use any function other than
the [Offset] function.
5.4.2.1 Description
•
P (Top of the screen)
Pressure value (unit: mmHg) for positive pressure sensor
•
Q (Top of the screen)
Pressure value (unit: mmHg) for supplementary pressure sensor
•
P (Bottom of the screen)
Output voltage value (unit: V) for positive pressure sensor
•
Q (Bottom of the screen)
Output voltage value (unit: V) for supplementary pressure sensor
•
[Offset]
Zero point correction for P and Q pressure values
5.4.2.2 Display
Summary of Contents for VaSera VS-1500
Page 12: ...Confidential Copying Prohibited x ...
Page 18: ...0 Chapter 1 Features 1 6 Confidential Copying Prohibited ...
Page 32: ...Chapter 4 Names of Parts of Equipment 4 6 Confidential Copying Prohibited ...
Page 42: ...Chapter 5 Maintenance 5 10 Confidential Copying Prohibited Diagonal Line Stripe ...
Page 178: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 3 ...
Page 180: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 5 ...
Page 191: ...Chapter 10 Part Location Diagram and Checkpoints 10 16 Confidential Copying Prohibited ...
Page 207: ...Chapter 11 Spare Parts List 11 16 Confidential Copying Prohibited ...