13. +12VA
Powers the analog circuits.
Allowable voltage is +12 ±0.6V.
Check voltage at the + side of C59.
(☞
10.1.1.B1
)
C59
+
12VA
14. +VISO
(+
5V
)
Powers the serial isolation unit.
Allowable voltage is +5.0 ±0.25V.
Check voltage at C131.
(☞
10.1.1.C2
)
+
VISO
(+
5V
)
15. +5VF
Powers the ECG isolation unit.
Allowable voltage is +5.0 ±0.25V.
Check voltage at the + side of C478.
(☞
10.1.1.B4
)
C4
7
8
+
5V
16. +2.5VF
Powers the ECG isolation unit.
Allowable voltage is +2.5 ±0.125V.
Check voltage at C467.
(☞
10.1.1.C4
)
+
2.5VF
C467
17. +5VPW
Powers the PCG/pulse wave isolation unit.
Allowable voltage is +5.0 ±0.25V.
Check voltage at C553.
(☞
10.1.1.A4
)
C
553
+
5.0V
18. P+2.5V
Powers the PCG/pulse wave isolation unit.
Allowable voltage is +2.5 ±0.125V.
Check voltage at C530.
(☞
10.1.1.A4
)
10-8
Confidential
Copying Prohibited
19. P
-
5V
Powers the PCG/pulse wave isolation unit.
Allowable voltage is -5.0 ±0.25V.
Check voltage at C541.
(☞
10.1.1.B4
)
C
530
+
5.0V
C541
P
-
5V
Summary of Contents for VaSera VS-1500
Page 12: ...Confidential Copying Prohibited x ...
Page 18: ...0 Chapter 1 Features 1 6 Confidential Copying Prohibited ...
Page 32: ...Chapter 4 Names of Parts of Equipment 4 6 Confidential Copying Prohibited ...
Page 42: ...Chapter 5 Maintenance 5 10 Confidential Copying Prohibited Diagonal Line Stripe ...
Page 178: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 3 ...
Page 180: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 5 ...
Page 191: ...Chapter 10 Part Location Diagram and Checkpoints 10 16 Confidential Copying Prohibited ...
Page 207: ...Chapter 11 Spare Parts List 11 16 Confidential Copying Prohibited ...