Chapter 8
Disassembly and Assembly
(8) Screw down the upper ECG amplifier cover SU.
Secure locations (2) marked
○
using:
W SEMS (L) M3 × 16
Secure location (1) marked
○
using:
P tight pan head WH 3 ×10
Make sure that D-Sub
connector lock fittings protrude
from locations marked
□
.
Upper ECG amplifier cover SU
First fit cover at location marked
○
.
(9) Attach the center shield plate and connect the BP cable.
Secure locations (5) marked
○
using:
W SEMS (L) M3 × 8
Connector locations (4)
Connect matching cable color with silk printing
Green (G), Black (B), Yellow (Y), Red (R)
* The yellow cable must be routed over the
green cable so that the latter does not come in
contact with the metal edge of the plate.
Attach and secure leftward from screw
locations
Tie base BS-20
Binder SKB-1M
8-14