Chapter 5
Maintenance
5-20
Confidential
Copying Prohibited
5.3.13 Testing PC Connection
Use this function to perform a ping test on the PC connected to the instrument to check if it is
properly connected.
5.3.13.1 Test Method
(1) Connect the VS-1500 to the network as described below.
(2) Press
the (Menu) key to open the [Top Menu], then touch [Initial Setting/Maintenance]
Æ
[Network and Master ID Setting]. Enter the IP address for the VS-1500 and VS-PC in the setup
window that appears.
(3) Press
the (Menu) key to open the [Top Menu], then touch [Initial Setting/Maintenance]
Æ
[Maintenance]
Æ
[OK (
↵
)]
Æ
[ping Test] to open the [ping Test] window.
(4) Touch [Start/End] to start the test. Touching [Start/End] during a test will cancel the test. Touching
[Start/End] again allows you to successively perform frequency and error count checks.
(5) Touch [Reset] to reset frequency and error count to "0".
(6) Touch [Return] to return to the [Maintenance menu].
5.3.13.2 Criteria
The error count after 100 tests must be "0".
5.3.13.3 Test Window
LAN cable (straight)
PC
Ethernet hub
Summary of Contents for VaSera VS-1500
Page 12: ...Confidential Copying Prohibited x ...
Page 18: ...0 Chapter 1 Features 1 6 Confidential Copying Prohibited ...
Page 32: ...Chapter 4 Names of Parts of Equipment 4 6 Confidential Copying Prohibited ...
Page 42: ...Chapter 5 Maintenance 5 10 Confidential Copying Prohibited Diagonal Line Stripe ...
Page 178: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 3 ...
Page 180: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 5 ...
Page 191: ...Chapter 10 Part Location Diagram and Checkpoints 10 16 Confidential Copying Prohibited ...
Page 207: ...Chapter 11 Spare Parts List 11 16 Confidential Copying Prohibited ...