Chapter 10 Part Location Diagram and Checkpoints
Confidential
Copying Prohibited
10-11
10.1.3.2 Fuse Check
1.
F1
F1
This fuse protects the +12 V power supply
providing power to the PCG/PW isolation unit.
(☞
10.1.1.B3
)
2.
F2
This fuse protects the +5 V power supply
providing power to the serial isolation unit.
F2
(☞
10.1.2.C3
)
3.
F3
This fuse protects the +5V power supply
providing power to the ECG isolation unit.
F3
(☞
10.1.2.B2
)
4.
F4
F4
This fuse protects the +5V power supply
providing power to externally connected serial devices.
(☞
10.1.2.C3
)
5.
F5
This fuse protects the ECG isolation unit power supply.
F5
(☞
10.1.2.B1
)