Chapter 7 Messages
7.1 Messages and Countermeasures
This chapter lists onscreen messages and provides information on the measures that should be
taken when they appear.
If the error message re-appears despite taking countermeasures, stop any
inspections and contact your Fukuda Denshi service representative.
7.1.1 Instrument and Measurement-related Messages
Message Cause Countermeasure
When the pulse wave was acquired,
the Cuff pressure was 5mmHg or
less.
Inspect the cuff and air hose for
damages or breakage.
The time to reach target-inflating
pressure was longer than specified.
・
If the cuffs start leaking air, press
the
(Start) key again to
measure.
・
Inspect the cuff and air hose for
damages or breakage.
AIR leak error
Loose “air” connector causes air
leakage.
Connect the air connectors
correctly.
BP error
Blood pressure measurement is not
possible.
Check whether cuffs are properly
set, have the patient relax and
re-measure.
BP low error
Because SYS is less than 30mmHg
and DIA is less than 20mmHg, the
blood pressure value cannot be
measured.
Confirmed that the cuff is
appropriately installed.
If the cuff is correctly attached, then
double check the pressure of the
patient with another system to
confirm low blood pressure.
BP re-inflation has been
turned off.
Repress off error
The re-inflation is turned off, and
could not inflate again when
re-inflation was necessary.
Set the setting of re-inflation to ON
or raise the pressure value.
Check PCG
The PCG microphone is not
connected.
Connect the PCG microphone.
Check PCG
II
sound.
The second heart sound cannot be
detected.
・
Check whether the ECG can be
correctly captured.
・
Adjust the PCG microphone
location.
Cuff error
Damaged cuff (rubber pouch)
causes air leakage.
Replace the cuff.
Cuff type error
When the inflation is intended for an
ankle cuff but a toe cuff is connected
or vice versa.
・
Connect a suitable cuff for the
measurement mode.
Illegal cartridge used.
A different type of cartridge is used. Make sure to use the same type of
cartridge as used before.
7-2
Confidential
Copying Prohibited
Summary of Contents for VaSera VS-1500
Page 12: ...Confidential Copying Prohibited x ...
Page 18: ...0 Chapter 1 Features 1 6 Confidential Copying Prohibited ...
Page 32: ...Chapter 4 Names of Parts of Equipment 4 6 Confidential Copying Prohibited ...
Page 42: ...Chapter 5 Maintenance 5 10 Confidential Copying Prohibited Diagonal Line Stripe ...
Page 178: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 3 ...
Page 180: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 5 ...
Page 191: ...Chapter 10 Part Location Diagram and Checkpoints 10 16 Confidential Copying Prohibited ...
Page 207: ...Chapter 11 Spare Parts List 11 16 Confidential Copying Prohibited ...