Chapter 9
Wiring Diagrams, Block Diagrams With Descriptions and Connector Pin Locations
9.3.2 Cuff Board /Pump Unit
9.3.2.1 Block Diagram
9.3.2.2 Description
Pressure Sensor (Main) Block
The input to the pressure sensor (main) is output to the Main Board as positive pressure P.
Positive pressure P goes through filters and an amplifier to produce pulse waves for CAVI
measurement (PW values) and pulse waves for BP measurement (OSC values). Main Board
control enables resetting of pulse waves.
Pressure Sensor (Supplementary) Block
The input to the pressure sensor (supplementary) is output to the Main Board as supplementary
pressure Q. The output from supplementary Q comparator is used in the BP safety circuit.
Pump
Unit
This unit uses a motor-driven pump to move air through the air filter to the pneumatic circuit to
raise air pressure. The flow rate valve and quick exhaust valves control the reduction of
pressure in the pneumatic circuit.
9-8
Confidential
Copying Prohibited
Summary of Contents for VaSera VS-1500
Page 12: ...Confidential Copying Prohibited x ...
Page 18: ...0 Chapter 1 Features 1 6 Confidential Copying Prohibited ...
Page 32: ...Chapter 4 Names of Parts of Equipment 4 6 Confidential Copying Prohibited ...
Page 42: ...Chapter 5 Maintenance 5 10 Confidential Copying Prohibited Diagonal Line Stripe ...
Page 178: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 3 ...
Page 180: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 5 ...
Page 191: ...Chapter 10 Part Location Diagram and Checkpoints 10 16 Confidential Copying Prohibited ...
Page 207: ...Chapter 11 Spare Parts List 11 16 Confidential Copying Prohibited ...