Chapter 5
Maintenance
Confidential
Copying Prohibited
5-21
5.3.14 Testing Operation Panel Fixed Keys
Use this function to test the fixed keys on the operation panel.
5.3.14.1 Test Method
(1) Press
the (Menu) key to open the [Top Menu], then touch [Initial Setting/Maintenance]
Æ
[Maintenance]
Æ
[OK (
↵
)]
Æ
[Fix Key Test] to open the [Fix Key Test] window.
(2) Press each fixed key on the operation panel to check that the corresponding symbol highlights on
the [Fix Key Test] window.
(3) Touch [Return] to return to the [Maintenance menu].
5.3.14.2 Criteria
The symbol corresponding to the pressed key must be highlighted.
5.3.14.3 Test Window