Instrument installation
Azbil Corporation
2-18
Model MTG11A/18A, MTG11B/18B, MTG14C
Installation on horizontal pipe
Required accessories
The following parts are required:
• Through-bolts and nuts
• Centering nuts
• Gaskets: The required gasket material will vary according to the material of the pipe
on which the detector is to be installed. See the installation procedures for
different pipe materials described on page 2-20 to page 2-25.
Procedure
Follow this procedure to install the detector on a horizontal pipe.
m
CAUTION
Improper installation may result in leakage or cause damage to the pipe flanges.
Step
Action
Drawing
1
• Insert through-bolts in the flange holes
shown by black dots in the drawing. Slip
two centering nuts onto each through-bolt
before inserting the bolts.
2
• Turn the detector so that the direction
mark on the detector matches the direc-
tion of fluid flow.
• Insert the detector and gaskets between
the pipe flanges.
• Position the detector so that it sits on top
of the centering.
3
• Make sure that the detector remains prop-
erly centered.
• Make sure that the gaskets do not pro-
trude beyond the edges of the pipe
flanges.
• When you have checked these items,
insert the remaining through-bolts into
the flange holes and tighten the bolts
evenly using the appropriate fastening
torque given on page 2-11.
Flange
Gasket
Direction of
fluid flow
CM2-MTG300-2001.book 18 ページ 2015年9月29日 火曜日 午前10時14分
Summary of Contents for MagneW Neo PLUS
Page 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Page 92: ...MEMO CM2 MTG300 2001 book 38...
Page 106: ...MEMO CM2 MTG300 2001 book 4...
Page 266: ...MEMO CM2 MTG300 2001 book 82...
Page 312: ...MEMO CM2 MTG300 2001 book 46...
Page 336: ...MEMO CM2 MTG300 2001 book 22...
Page 338: ...CM2 MTG300 2001 book 2...