![Azbil MagneW Neo PLUS User Manual Download Page 86](http://html1.mh-extra.com/html/azbil/magnew-neo-plus/magnew-neo-plus_user-manual_3042666086.webp)
Instrument installation
Azbil Corporation
2-32
Model MTG11A/18A, MTG11B/18B, MTG14C
Installation on metal pipe (2)
Introduction
The installation method described in this section is to be used with the following
grounding ring materials. For the installation method used with grounding rings of
SUS material, refer to the table on page 2-30.
Pipe material: metal
Grounding ring material: Non-SUS material
Required accessories
The following parts are required. No gaskets are necessary since PTFE gaskets are
provided.
• Bolts and nuts
Installation procedure
• Install the detector as shown in Figure 2-28. See Table 2-5 on page 2-27 for the
appropriate fastening torque.
m
CAUTION
Please note that the use of an additional gasket besides the existing PTFE gasket
may result in leakage (see Figure 2-29).
Figure 2-28 Installation using Non-SUS material grounding ring
Figure 2-29 Example of incorrect installation
Lining
Grounding ring
PTFE gasket
(supplied)
Gasket
Lining
Grounding ring
PTFE gasket
(supplied)