Azbil Corporation
Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C
6-67
6
or
Press the
key and enter into
the CORRECT DAC screen.
Current value shown in the bottom
right screen presents the
adjustment variation.
By pressing the
key, you can
change from 0.002, 0.01, 0.05 to
0.25 mA.
Adjust by pressing the
key or
key.
7
Press the
key.
Display return back to the screen
shown to the right. # will be shown
in the bottom right screen. This
shows that it is still in constant
current generation mode. You need
to clear this mode.
8
Press the
key again.
9
Press the
key.
Confirm that the “#” mark has been
deleted.
10
Adjust to 100% (20 mA) by
following the same procedure as
for adjusting 0% (4 mA).
Step
Key
Procedure
SFC screen
ENTER
( Yes )
NON-VOL
MENU I
ITEM
DE CONF.
H
NEXT
L
PREV
ENTER
( Yes )
NON-VOL
MENU I
ITEM
DE CONF.
H
NEXT
L
PREV
SFCM00006121D
CLR
( No )
CLR
( No )
SFCM00006119D
OUT- J
PUT
INPUT
OUT- J
PUT
INPUT
SFCM00006117D
CLR
( No )
CLR
( No )
SFCM00006009D
CM2-MTG300-2001.book 67 ページ 2015年9月29日 火曜日 午前10時14分
Summary of Contents for MagneW Neo PLUS
Page 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Page 92: ...MEMO CM2 MTG300 2001 book 38...
Page 106: ...MEMO CM2 MTG300 2001 book 4...
Page 266: ...MEMO CM2 MTG300 2001 book 82...
Page 312: ...MEMO CM2 MTG300 2001 book 46...
Page 336: ...MEMO CM2 MTG300 2001 book 22...
Page 338: ...CM2 MTG300 2001 book 2...