Operation using the data setting device
Azbil Corporation
6-48
Model MTG11A/18A, MTG11B/18B, MTG14C
Setting detector constant: [EX(mA)] function
Introduction
The detector constant of the converter is set at the factory according to the ordered
specifications. This constant can be changed.
Used when
When a combination between the detector and converter has been changed, the
detector constant set by the converter needs to be changed.
Procedure
Use the following procedure to set detector constant.
Step
Key
Procedure
SFC screen
1
Make sure that the SFC is set to
“READY”.
If it is not, press the
key to set it
to “READY”.
2
Press the
key to access the
CONFIG functions.
3
or
Press the
key or
key to
display the screen as shown on the
right.
4
Press the
key.
Result:
• The currently set detector constant
appears as shown on the right.
5
Press numeric keys to set the detector
constant. Setting range is 200 to
699.9.
CLR
( No )
CLR
( No )
SFCM00006009D
B
CONF
B
CONF
SFCM00006006D
SFCM00006018D
H
NEXT
L
PREV
H
NEXT
L
PREV
SFCM00006019D
ENTER
( Yes )
NON-VOL
ENTER
( Yes )
NON-VOL
SFCM00006020D
SFCM00006021D
SFCM00006022D
CM2-MTG300-2001.book 48 ページ 2015年9月29日 火曜日 午前10時14分
Summary of Contents for MagneW Neo PLUS
Page 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Page 92: ...MEMO CM2 MTG300 2001 book 38...
Page 106: ...MEMO CM2 MTG300 2001 book 4...
Page 266: ...MEMO CM2 MTG300 2001 book 82...
Page 312: ...MEMO CM2 MTG300 2001 book 46...
Page 336: ...MEMO CM2 MTG300 2001 book 22...
Page 338: ...CM2 MTG300 2001 book 2...