Operation using the data setting device
Azbil Corporation
5-64
Model MTG11A/18A, MTG11B/18B, MTG14C
5-7-2 : Performing Loop Checks of Analog Outputs by using a
calibrator Model MGZ.
Analog Output Check
With the signal input by a calibrator Model MGZ, the electromagnetic flowmeter
outputs analog 4 to 20mA to perform the loop check. Other instruments in the analog
current output loop, such as recorders and controllers can be checked.
Default setting
Signal input by the calibrator Model MGZ
Setting range
0%, 25% to 100%
(The % flow rate input from 1% to 24% are not available.)
Perform loop checks of analog outputs by using a calibrator Model MGZ in
accordance with the following procedures.:
Step
Procedure
Screen
1
Enter MAINTENANCE MODE (see “Entering
Engineering Mode and Maintenance Mode” on
page 5-36
). Then the screen at right is displayed.
2
Press the key to move the cursor to the OFF
position.
Press the key. With the displays switched from
OFF to ON, the output check mode is then active.
3
Input the signal from the calibrator and perform
the loop check.
For the details of the calibrator operation, refer to
the users manual of the MGZ calibrator.
20.0 %
_> OUTPUT CHECK
MODE OFF
20.0 %
_> OUTPUT CHECK
MODE _
OFF
20.0 %
_> OUTPUT CHECK
MODE _
ON
20.0 %
_> OUTPUT CHECK
W/CALIBRATOR
20.0 %
_> OUTPUT CHECK
W/CALIBRATOR
CM2-MTG300-2001.book 64 ページ 2015年9月29日 火曜日 午前10時14分
Summary of Contents for MagneW Neo PLUS
Page 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Page 92: ...MEMO CM2 MTG300 2001 book 38...
Page 106: ...MEMO CM2 MTG300 2001 book 4...
Page 266: ...MEMO CM2 MTG300 2001 book 82...
Page 312: ...MEMO CM2 MTG300 2001 book 46...
Page 336: ...MEMO CM2 MTG300 2001 book 22...
Page 338: ...CM2 MTG300 2001 book 2...