Instrument installation
Azbil Corporation
2-4
Model MTG11A/18A, MTG11B/18B, MTG14C
Criteria for installation location (2)
Installed position
Install the instrument at a place where the measured fluid always fills inside the
detector. An example of installation is shown in the figure below to illustrate this
condition.
~Note
•
The detector must be positioned as shown by the circled areas in the
figure shown above. If the pipe is not filled output errors will occur.
•
If the measured fluid is highly viscous, we recommend installing the
instrument on a vertical pipe, in order to secure an axial symmetrical
flow.
•
Install a straight pipe section on the upstream side of the detector. For
the straight section length, see the figure below.
Figure 2-2 Straight pipe section upstream of detector (D: Nominal detector bore
diameter)
Figure 2-1 Example of installation
The fluid
may not fill
completely.
The fluid
may not fill
completely.
Prone to air lock
Pump
Good
Good
Diverging pipe section
with a cone angle > 15
(< 15 considered as
straight pipe
section)
90-degree bend
Detector
Detector
Detector
Detector
Detector
Detector
5D or over
5D or over
5D or over
10D or over
10D or over
10D or over
5D or over
Pump of various types
T
Gate valve
fully
open
Valve of various types
Converging pipe section
(considered as
straight pipe
section)
Upstream
CM2-MTG300-2001.book 4 ページ 2015年9月29日 火曜日 午前10時14分
Summary of Contents for MagneW Neo PLUS
Page 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Page 92: ...MEMO CM2 MTG300 2001 book 38...
Page 106: ...MEMO CM2 MTG300 2001 book 4...
Page 266: ...MEMO CM2 MTG300 2001 book 82...
Page 312: ...MEMO CM2 MTG300 2001 book 46...
Page 336: ...MEMO CM2 MTG300 2001 book 22...
Page 338: ...CM2 MTG300 2001 book 2...