Azbil Corporation
Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C
5-45
5-6-4 : Setting detector factor
Set the detector factor. The C1 value is always 1.0000. Set the C2 value to that shown
on the detector nameplate under the heading Detector Factor.
Step
Procedure
Screen
1
Enter ENGINEERING MODE (see “Entering
Engineering Mode and Maintenance Mode” on
page 5-36
). Then press the or key to display
the screen at right.
2
Press the key to set the detector factor.
C1 is always set to 1.0000.
Set the C2 factor to that stamped on the nameplate
of the flowmeter.
3
Press the key to move the cursor to the position
under #.
Press the MODE key to return to the
MEASURING MODE and to save the data.
20.0 %
_# C1 1.0000
C2 1.0234
20.0 %
_# C1 1.0000
C2 1.0234
20.0 %
_# C1 1.0000
C2 1.0234