Operation using the data setting device
Azbil Corporation
5-34
Model MTG11A/18A, MTG11B/18B, MTG14C
5-5-9 : Selecting a communication system
Select the communication system (SFC, DE, HART, and communication disable).
Select a communication system to be used. Note that the converter will be rebooted
when the MODE key is pressed to switch to the MEASURING MODE after the
settings are changed.
HART: HART communication by using the HART Communicator.
SFN.A: SFC in the analog (4-20 mA) output mode.
SFN.D: DE (Digital Enhanced) communication.
NONE: Communication is not used/disable.
By default, SFN.A: SFC communication is set.
Select a communication system in accordance with the following procedure:
Step
Procedure
Screen
1
The screen at right is a screen display example in
MEASURING MODE (measurement state).
Press the MODE key.
2
The OPERATOR'S MODE screen appears for
approx. two seconds and then the damping setting
screen appears.
3
Press the key three times to display the screen as
shown on the right.
4
Press the key to move the cursor to the position
for the communication system (SFN.A, SFN.D,
NONE, HART). The screen at right shows an
example where SFN.A has been selected for the
communication system.
5
Press the or key to select a desired
communication system to be set.
The screen at right shows an example where
HART communication has been selected.
20.0 %
01.94 m
3
/h
WP0
00069401
20.0 %
* OPERATOR'S
MODE
20.0 %
* DAMPING
005.0 s
20.0 %
* COM SELECT
SFN. A
20.0 %
* COM SELECT
SFN. A
20.0 %
* COM SELECT
HART
CM2-MTG300-2001.book 34 ページ 2015年9月29日 火曜日 午前10時14分
Summary of Contents for MagneW Neo PLUS
Page 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Page 92: ...MEMO CM2 MTG300 2001 book 38...
Page 106: ...MEMO CM2 MTG300 2001 book 4...
Page 266: ...MEMO CM2 MTG300 2001 book 82...
Page 312: ...MEMO CM2 MTG300 2001 book 46...
Page 336: ...MEMO CM2 MTG300 2001 book 22...
Page 338: ...CM2 MTG300 2001 book 2...