Operation using the data setting device
Azbil Corporation
6-60
Model MTG11A/18A, MTG11B/18B, MTG14C
Select pulse output / contact output [DIGITAL I/O] function
Two wired magnetic flowmeter can be configured for pulse output or contact output
other than the analog current output (4-20 mA).
Use the following procedure to configure the pulse output and contact output.
Step
Key
Procedure
SFC screen
1
Confirm that the SFC is set to
“READY”.
If it is not, press the
key to
set it to “READY”.
2
or
Press the
key to access the
CONFIG functions. Press the
key or
key to display the
screen as shown on the right
3
or
Press the
key to access the
DIGITAL I/O function, press the
key or
key to display
the screen as shown on the right.
Present setting function appears.
DO NOT USED..... Pulse output
HI-LO ALARM..... Contact output
* Following function will appear
on the screen but are not available
for two wired magnetic flowmeter.
ALARM
DIAG
EMPTY
H1-L1/H2-L2 ALM
or
4
Press the
key and specify the
function to be set.
5
Press the
key.
The changed setting is saved to the
SFC.
CLR
( No )
CLR
( No )
SFCM00006009D
B
CONF
H
NEXT
L
PREV
B
CONF
H
NEXT
L
PREV
SFCM00006100D
ENTER
( Yes )
NON-VOL
H
NEXT
L
PREV
ENTER
( Yes )
NON-VOL
H
NEXT
L
PREV
SFCM00006101D
SFCM00006102D
MENU I
ITEM
DE CONF.
MENU I
ITEM
DE CONF.
SFCM00006102D
ENTER
( Yes )
NON-VOL
ENTER
( Yes )
NON-VOL
SFCM00006103D
CM2-MTG300-2001.book 60 ページ 2015年9月29日 火曜日 午前10時14分
Summary of Contents for MagneW Neo PLUS
Page 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Page 92: ...MEMO CM2 MTG300 2001 book 38...
Page 106: ...MEMO CM2 MTG300 2001 book 4...
Page 266: ...MEMO CM2 MTG300 2001 book 82...
Page 312: ...MEMO CM2 MTG300 2001 book 46...
Page 336: ...MEMO CM2 MTG300 2001 book 22...
Page 338: ...CM2 MTG300 2001 book 2...