Azbil Corporation
Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C
5-59
5-6-14 : Selecting failsafe mode for pulse output
This function is used to determine the pulse output direction when the flowmeter
detects a critical status condition.
Set range:
OFF
Outputs no pulse.
HOLD Pulse output signal held at its present state
Default:
OFF
Set failsafe mode for pulse output in accordance with the following procedure:
m
CAUTION
The failsafe mode is very important for the overall safety of the control process. Choose the
failsafe direction carefully, as equipment damage can result from a wrong choice.
Step
Procedure
Screen
1
Enter ENGINEERING MODE in accordance with
the entry into ENGINEERING MODE (see
section 5-5-9 on page 5-36). Then press the or
key to display the screen at right.
2
Press the key.
3
Using the or key, determine the failsafe mode
for pulse output.
4
Press the key to move the cursor to the position
under #.
Press the MODE key to return to the
MEASURING MODE and to save data.
m
CAUTION
You have only ten minutes to return to MEASURING MODE to save the new value before
the system resets it to the previously saved value.
12.3 %
# ERROR OUT MODE
P.OUT OFF
12.3 %
# ERROR OUT MODE
P.OUT OFF
12.3 %
# ERROR OUT MODE
P.OUT HOLD
12.3 %
# ERROR OUT MODE
P.OUT HOLD
CM2-MTG300-2001.book 59 ページ 2015年9月29日 火曜日 午前10時14分
Summary of Contents for MagneW Neo PLUS
Page 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Page 92: ...MEMO CM2 MTG300 2001 book 38...
Page 106: ...MEMO CM2 MTG300 2001 book 4...
Page 266: ...MEMO CM2 MTG300 2001 book 82...
Page 312: ...MEMO CM2 MTG300 2001 book 46...
Page 336: ...MEMO CM2 MTG300 2001 book 22...
Page 338: ...CM2 MTG300 2001 book 2...