Operation using HART communicator
Azbil Corporation
7-38
Model MTG11A/18A, MTG11B/18B, MTG14C
Pulse output check
To output pulse fixed value, proceed as follow:
4.
Use the numeric keys to enter a fixed
value to be output in the value input
display. Setting range is duty ratio of
0% to +100%. When value is entered,
press F4 (ENTER). Analog current
corresponding to its output will be
output. Press F3 (ABORT) to return to
the previous menu.
Figure 7-82
1.
From online menu,
Select:
1. Device setup
2. Diag/Service
2. Loop test
3. Out put check Pls
Figure 7-83 will then be displayed.
If the control system is not affected by
the forcible change of the current
signal, press F4(OK). By pressing
F3(ABORT), the procedure is
canceled.
Figure 7-83
2.
Figure 7-84 will then appear. Choose
START to start fixed value of pulse
output, and then press F4(ENTER)
Figure 7-84
3.
Note of caution will appear as shown
in Figure 7-85. To proceed, press
F4(OK).
Figure 7-85
CM2-MTG300-2001.book 38 ページ 2015年9月29日 火曜日 午前10時14分