Azbil Corporation
Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C
6-77
5
Press the
key and select the
pulse width to be set. (Select from
duty 50, 1/7/10/15/30/50/100/200
ms)
You can enter it directly by
pressing the numeric keys. Range
of two wired magnetic flowmeter
is 1 to 1000 ms, however, SFC can
only set up to 999.9 ms.
6
Press the
key.
The changed setting is saved on the
SFC.
7
or
Press the
key or
key to
display the screen as shown on the
right.
8
Press the
key.
Changed setting data is written into
the database of the converter.
9
Press the
key and the screen
returns to the screen as in step 1.
Step
Key
Procedure
SFC screen
MENU I
ITEM
DE CONF.
MENU I
ITEM
DE CONF.
SFCM00006151D
ENTER
( Yes )
NON-VOL
ENTER
( Yes )
NON-VOL
SFCM00006145D
H
NEXT
L
PREV
H
NEXT
L
PREV
SFCM00006152D
ENTER
( Yes )
NON-VOL
ENTER
( Yes )
NON-VOL
SFCM00006149D
CLR
( No )
CLR
( No )
SFCM00006009D