Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C
5-57
5-6-12 : Setting upper and lower limit alarm
This function is used to set the upper and lower limit alarm set points when the contact
output is selected.
An alarm is output when the flow rate exceeds these preset upper and lower limits.
The alarm output status depends on the “Setting contact output status” described later.
Set range:
HI-ALM
0% to +115%
LO-ALM
0% to +115%
Default:
HI-ALM
+115%
LO-ALM
0%
Set the upper/lower limit alarm in accordance with the following procedure:
However, set as follows: HI-ALM > LO-ALM.
Step
Procedure
Screen
1
Enter ENGINEERING MODE in accordance with
the entry into ENGINEERING MODE (see
section 5-5-9 on page 5-36). Then press the or
key to display the screen at right.
2
Using the key, move the cursor to the position
under a digit to be set or changed.
3
Using the or key, change the value to the
desired value to be set.
4
Press the key to move the cursor to the position
under #.
Press MODE key to return to the MEASURING
MODE and to save data.
12.3 %
# HI-ALM
+115%
LO-ALM +000%
12.3 %
# HI-ALM
+100%
LO-ALM -000%
12.3 %
# HI-ALM
+080%
LO-ALM -000%
12.3 %
# HI-ALM
+080%
LO-ALM -000%
CM2-MTG300-2001.book 57 ページ 2015年9月29日 火曜日 午前10時14分
Summary of Contents for MagneW Neo PLUS
Page 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Page 92: ...MEMO CM2 MTG300 2001 book 38...
Page 106: ...MEMO CM2 MTG300 2001 book 4...
Page 266: ...MEMO CM2 MTG300 2001 book 82...
Page 312: ...MEMO CM2 MTG300 2001 book 46...
Page 336: ...MEMO CM2 MTG300 2001 book 22...
Page 338: ...CM2 MTG300 2001 book 2...