Operation using the data setting device
Azbil Corporation
5-26
Model MTG11A/18A, MTG11B/18B, MTG14C
(To the next page)
Step
Procedure
Screen
9
Next, fill fluid into the pipe. If the scale appears
on the electrode, clean the electrode and fill fluid.
Check that the empty or scale on electrode status
is not detected in this condition.
Since it takes at least 30 seconds to clear the
empty or scale on electrode status after fluid is
filled in the pipe, perform checking when 30
seconds or more have passed after fluid is filled.
(Branch 1)
• When the result is (1) in step 8
Check that the empty or scale on electrode status
is not detected when SENSITIVITY LLL has
been set.
(Result)
• If the empty or scale on electrode status is not
detected, use that SENSITIVITY LLL setting
without change.
• If the empty or scale on electrode status is
detected and the screen shown on the right
appears, this function is not available in that
environment. Set the function to OFF.
(Branch 2)
• When the result is (2) in step 8
Check that the empty or scale on electrode status
is not detected when SENSITIVITY LL has
been set.
(Result)
• If the empty or scale on electrode status is not
detected, use that SENSITIVITY LL setting
without change.
• If the empty or scale on electrode status is
detected and the screen shown on the right
appears, this function is not available in that
environment. Set the function to OFF.
(Branch 3)
• When the result is (3) in step 8
Check that the empty or scale on electrode status
is not detected when SENSITIVITY LOW has
been set.
(Result)
• If the empty or scale on electrode status is not
detected, use that SENSITIVITY LOW setting
without change.
• If the empty or scale on electrode status is
detected and the screen shown on the right
appears, this function is not available in that
environment. Set the function to OFF.
0.0 %
EMPTY OR SCALE
ON ELECTRODE
0.0 %
EMPTY OR SCALE
ON ELECTRODE
0.0 %
EMPTY OR SCALE
ON ELECTRODE
CM2-MTG300-2001.book 26 ページ 2015年9月29日 火曜日 午前10時14分
Summary of Contents for MagneW Neo PLUS
Page 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Page 92: ...MEMO CM2 MTG300 2001 book 38...
Page 106: ...MEMO CM2 MTG300 2001 book 4...
Page 266: ...MEMO CM2 MTG300 2001 book 82...
Page 312: ...MEMO CM2 MTG300 2001 book 46...
Page 336: ...MEMO CM2 MTG300 2001 book 22...
Page 338: ...CM2 MTG300 2001 book 2...