Operation using the data setting device
Azbil Corporation
5-28
Model MTG11A/18A, MTG11B/18B, MTG14C
Electrode status diagnostic flow chart
Install the MTG 2 to the pipe. Pipe should be empty
Electrode Status Diagnostic function
Default setting: SENSITIVITY OFF
Change the SENSITIVITY to HIGH
when the pipe is empty or the scale appears on electrode.
Electrode Status Diagnostic function is not applicable.
Change the SENSITIVITY to OFF.
Blink
"EMPTY OR SCALE
ON ELECTRODE"
Blink
"EMPTY OR SCALE
ON ELECTRODE"
Change the SENSITIVITY to
HIGH when the flowtube is
filled with the process fluid.
Blink
"EMPTY OR SCALE
ON ELECTRODE"
Use the Electrode Status
Diagnostic function with the
SENSITIVITY HIGH.
Change the SENSITIVITY to
MID when the flowtube is filled
with the process fluid.
Use the Electrode Status
Diagnostic function with the
SENSITIVITY MID.
Blink
"EMPTY OR SCALE
ON ELECTRODE"
Change the SENSITIVITY to LOW
when the pipe is empty or the scale
appears on electrode.
Blink
"EMPTY OR SCALE
ON ELECTRODE"
Change the SENSITIVITY to LL
when the pipe is empty or the scale
appears on electrode.
Electrode Status Diagnostic
function is not applicable.
Change the SENSITIVITY
to OFF.
Electrode Status Diagnostic
function is not applicable.
Change the SENSITIVITY
to OFF.
Change the SENSITIVITY to MID
when the pipe is empty or the scale appears on electrode.
NO
YES
NO
NO
YES
NO
YES
YES
NO
YES
CM2-MTG300-2001.book 28 ページ 2015年9月29日 火曜日 午前10時14分
Summary of Contents for MagneW Neo PLUS
Page 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Page 92: ...MEMO CM2 MTG300 2001 book 38...
Page 106: ...MEMO CM2 MTG300 2001 book 4...
Page 266: ...MEMO CM2 MTG300 2001 book 82...
Page 312: ...MEMO CM2 MTG300 2001 book 46...
Page 336: ...MEMO CM2 MTG300 2001 book 22...
Page 338: ...CM2 MTG300 2001 book 2...