Azbil Corporation
Instrument installation
Model MTG11A/18A, MTG11B/18B, MTG14C
2-21
Installation on metal pipe (2)
Introduction
The installation method described in this section corresponds to the following
combination of pipe and grounding ring materials. For the installation method
corresponding to any other combination, refer to the table on page 2-17.
Pipe material: metal
Grounding ring material: Non-SUS
Required accessories
The following parts are required. No gaskets are necessary since PTFE gaskets are
provided.
• Through-bolts and nuts
• Centering nuts
Installation procedure
• Install the detector as shown following figures. See Table 2-1 on page 2-12 for the
appropriate fastening torque.
• To use rubber gaskets for a low fastening torque, refer to page 2-25.
m
CAUTION
Please note that the use of an additional gasket besides the existing PTFE gasket
may result in leakage (see Figure 2-17)
Figure 2-16 Installation using Non-SUS material grounding ring and metal pipe
Figure 2-17 Example of incorrect installation
PTFE gasket
Grounding ring
Lining
PTFE gasket
Grounding ring
Lining
Gasket
CM2-MTG300-2001.book 21 ページ 2015年9月29日 火曜日 午前10時14分
Summary of Contents for MagneW Neo PLUS
Page 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Page 92: ...MEMO CM2 MTG300 2001 book 38...
Page 106: ...MEMO CM2 MTG300 2001 book 4...
Page 266: ...MEMO CM2 MTG300 2001 book 82...
Page 312: ...MEMO CM2 MTG300 2001 book 46...
Page 336: ...MEMO CM2 MTG300 2001 book 22...
Page 338: ...CM2 MTG300 2001 book 2...