Operation using the data setting device
Azbil Corporation
6-68
Model MTG11A/18A, MTG11B/18B, MTG14C
Calibrating gain constant [GAIN CAL] function
Gain constant of amplifier which is set inside the converter can be calibrated by using
SFC. To do this, Azbil Corporation’s smart calibrator (model MGZ14) will be
required.
Use the following procedure to calibrate the gain constant.
Step
Key
Procedure
SFC screen
1
Confirm that the SFC is set to
“READY”.
If it is not, press the
key to
set it to “READY”.
2
or
Press the
key to access the
CONFIG functions. Press the
key or
key to display the
screen as shown on the right
3
or
Press the
key. Press the
key or
key to display the
screen as shown on the right
4
Press the
key.
For gain calibration press the
key, again.
Connect smart calibrator model
MGZ14 to the converter; (For
connecting and operating smart
calibrator model MGZ14, refer to
its user’s manual CM2-MGZ200-
2001).
CLR
( No )
CLR
( No )
SFCM00006009D
B
CONF
H
NEXT
L
PREV
B
CONF
H
NEXT
L
PREV
SFCM00006020D
SFCM00006122D
ENTER
( Yes )
NON-VOL
H
NEXT
L
PREV
ENTER
( Yes )
NON-VOL
H
NEXT
L
PREV
SFCM00006123D
ENTER
( Yes )
NON-VOL
ENTER
( Yes )
NON-VOL
ENTER
( Yes )
NON-VOL
SFCM00006124D
CM2-MTG300-2001.book 68 ページ 2015年9月29日 火曜日 午前10時14分
Summary of Contents for MagneW Neo PLUS
Page 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Page 92: ...MEMO CM2 MTG300 2001 book 38...
Page 106: ...MEMO CM2 MTG300 2001 book 4...
Page 266: ...MEMO CM2 MTG300 2001 book 82...
Page 312: ...MEMO CM2 MTG300 2001 book 46...
Page 336: ...MEMO CM2 MTG300 2001 book 22...
Page 338: ...CM2 MTG300 2001 book 2...