Operation using the data setting device
Azbil Corporation
6-80
Model MTG11A/18A, MTG11B/18B, MTG14C
Setting counter reset function [RESET TOTALZE] function
Resetting the totalized value and setting reset value are possible using the SFC. When
resetting the totalized value, two wired magnetic flowmeter internal counter will be
reset and start resetting from the reset value which has been set. You can enter the reset
value maximum of 10 figures to SFC, however for two wired magnetic flowmeter only
8 figures are effective. Use the following procedure to reset the totalized value and to
set the reset value.
Step
Key
Procedure
SFC screen
1
Confirm that the SFC is set to
“READY”.
If it is not, press the
key to
set it to “READY”.
2
or
Press the
key to access the
CONFIG functions. Press the
key or
key to display the
screen as shown on the right.
3
or
Press the
key to access the
TOTALIZER functions. Press the
key or
key to display
the screen as shown on the right.
4
Press the
key. Display the
actual reset value.
5
Enter the reset value to be set
pressing the numeric keys.
CLR
( No )
CLR
( No )
SFCM00006009D
B
CONF
H
NEXT
L
PREV
B
CONF
H
NEXT
L
PREV
SFCM00006020D
SFCM00006136D
ENTER
( Yes )
NON-VOL
H
NEXT
L
PREV
ENTER
( Yes )
NON-VOL
H
NEXT
L
PREV
SFCM00006155D
ENTER
( Yes )
NON-VOL
ENTER
( Yes )
NON-VOL
SFCM00006156D
SFCM00006157D
CM2-MTG300-2001.book 80 ページ 2015年9月29日 火曜日 午前10時14分
Summary of Contents for MagneW Neo PLUS
Page 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Page 92: ...MEMO CM2 MTG300 2001 book 38...
Page 106: ...MEMO CM2 MTG300 2001 book 4...
Page 266: ...MEMO CM2 MTG300 2001 book 82...
Page 312: ...MEMO CM2 MTG300 2001 book 46...
Page 336: ...MEMO CM2 MTG300 2001 book 22...
Page 338: ...CM2 MTG300 2001 book 2...