Operation using the data setting device
Azbil Corporation
6-44
Model MTG11A/18A, MTG11B/18B, MTG14C
Setting or changing low flow cutoff: [CUT-OFF] function
Introduction
When a fluid in the detector is flowing extremely slowly, the converter judges that the
fluid is stationary and outputs a signal (4 mA DC in case of analog output) equivalent
to a flow rate of zero. The value, which becomes the threshold of this judgment is
called “low flow cutoff”.
The low flow cutoff is set using a percentage over the upper range value of the flow
rate measurement range set by the URV.
Procedure
Use the following procedure to set low flow cutoff.
m
CAUTION
The low flow cutoff is a factor of extreme importance that affects the operation of the entire
control process. Define the range to be controlled and start the setting carefully.
Step
Key
Procedure
SFC screen
1
Confirm that the SFC is set to
“READY”.
If it is not, press the
key to set
it to “READY”.
2
Press the
key to access the
CONFIG functions.
3
Press the
key.
4
Press the
key.
5
Pressing the
key changes a
numeric value to the right of CUT
OFF=ON displayed on the screen
from 0% up to 10% in 1 point
increments. Continuing to press the
key further sets CUT
OFF=OFF and displays CUT
OFF=ON 0% again. Display the low
flow cutoff to be set in the range of
1- to 10% on the screen.
CLR
( No )
CLR
( No )
SFCM00006009D
B
CONF
B
CONF
SFCM00006018D
ENTER
( Yes )
NON-VOL
ENTER
( Yes )
NON-VOL
SFCM00006020D
H
NEXT
H
NEXT
MENU I
ITEM
DE CONF.
MENU I
ITEM
DE CONF.
MENU I
ITEM
DE CONF.
MENU I
ITEM
DE CONF.
SFCM00006059D
CM2-MTG300-2001.book 44 ページ 2015年9月29日 火曜日 午前10時14分
Summary of Contents for MagneW Neo PLUS
Page 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Page 92: ...MEMO CM2 MTG300 2001 book 38...
Page 106: ...MEMO CM2 MTG300 2001 book 4...
Page 266: ...MEMO CM2 MTG300 2001 book 82...
Page 312: ...MEMO CM2 MTG300 2001 book 46...
Page 336: ...MEMO CM2 MTG300 2001 book 22...
Page 338: ...CM2 MTG300 2001 book 2...