Azbil Corporation
Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C
6-29
Making zero adjustment: CORRECT key
Use the following procedure to do the auto zero adjustment from the SFC.
When adjusting auto zero point, stop and make static the fluid in the flow meter.
Step
Key
Procedure
SFC screen
1
Make sure that the SFC is set to
“READY”.
If a screen other than the one here to the
right is displayed, press the
key.
2
Press the
key, and then press the
key,
Result:
The actual measured value appears as
shown on the right.
3
Press the
key. ‘Zero INPUT?’ will
be displayed. Confirm that the actual
flow rate equals to zero.
4
After checking the flow rate, press the
key.
Result:
Auto zero point adjustment is initiated.
It takes approx. 2 min. When the screen
returns to the previous screen, auto zero
adjustment is completed.
5
Press the
key to return to at step 1.
CLR
( No )
CLR
( No )
SFCM00006009D
SHIFT
OUT- J
PUT
INPUT
SHIFT
OUT- J
PUT
INPUT
SFCM00006075D
COR- K
RECT
RESET
COR- K
RECT
RESET
SFCM00006076D
ENTER
( Yes )
NON-VOL
ENTER
( Yes )
NON-VOL
SFCM00006077D
SFCM00006075D
CLR
( No )
CLR
( No )
SFCM00006078D
CM2-MTG300-2001.book 29 ページ 2015年9月29日 火曜日 午前10時14分
Summary of Contents for MagneW Neo PLUS
Page 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Page 92: ...MEMO CM2 MTG300 2001 book 38...
Page 106: ...MEMO CM2 MTG300 2001 book 4...
Page 266: ...MEMO CM2 MTG300 2001 book 82...
Page 312: ...MEMO CM2 MTG300 2001 book 46...
Page 336: ...MEMO CM2 MTG300 2001 book 22...
Page 338: ...CM2 MTG300 2001 book 2...