Instrument installation
Azbil Corporation
2-10
Model MTG11A/18A, MTG11B/18B, MTG14C
Wiring connection distance of remote-type converter and detector
When installing the converter and the detector, there is limitation on the cable length
by the conductivity of the fluid to be measured and the diameter of the detector. Select
the wiring distance (cable length) after confirming the following specification. Use our
special cable (model SMC11) for the wiring between the converter and the detector.
For the diameter 10 mm or 15 mm
For the diameter 25 mm or more
Figure 2-7 Cable length (m) between converter and detector
Figure 2-8 Cable length (m) between converter and detector
30
2
50
C
onduc
tivity (ms/cm)
Cable length (m)
70
2
10
C
onduc
tivity (ms/cm)
Cable length (m)