IOM

-

SENTRY

-

Ex

-

C: MAY 2020

 

 

Installation, Operation & 

 

Maintenance 

Instructions

 

Sentry Series

 

Models P01, P02, P03 (Pressure Switches)

 

Models D01, D02, D03 (Differential Pressure 

Switches)

 

Models T01, T02 (Temperature Switches)

 

General

 

The  unit  is  manufactured,  checked  and  supplied  in 

accordance  with  our  published  specification,  and 

when  installed  and  used  in  normal  or  prescribed 

applications,  with  the  lid  in  place  and  within  the 

parameters  set  for  mechanical  and  electrical 

performance, will not cause danger or hazard to life 

or limb.

 

 

THE  USERS  ATTENTION  IS  DRAWN  TO 

THE  FACT  THAT,  WHEN  THE  UNIT  IS 

LIVE

’ 

WITH  RESPECT  TO  ELECTRICAL 

OR  PRESSURE  SUPPLIES,  A  HAZARD 

MAY  EXIST  IF  THE  UNIT  IS  OPENED  OR 

DISMANTLED.

 

 

UNITS  MUST  BE  SELECTED  AND 

INSTALLED  BY  SUITABLY  TRAINED  AND 

QUALIFIED 

PERSONNEL 

IN 

ACCORDANCE 

WITH 

APPROPRIATE 

CODES  OF  PRACTICE  SO  THAT  THE 

POSSIBILITY OF FAILURE RESULTING IN 

INJURY  OR  DAMAGE  CAUSED  BY 

MISUSE 

OR 

MISAPPLICATION 

IS 

AVOIDED.

 

 

Operating Principles

 

Pressure  Switch  models  P0*  and  Differential 

Pressure  Switch  models  D0*  are  diaphragm 

operated switches.

 

 

These  diaphragms  generate  a  force  proportional  to 

the  applied  pressure  and  are  balanced  by  a  user 

adjustable  control  spring.  When  the  force  exceeds 

that  created  by  the  control  spring,  the  diaphragm 

moves causing a push rod to actuate a snap

-

acting 

micros

-

witch. 

 

 

Temperature  models  T01  &  T02  work  in  the  same 

way as the Pressure models with the exception that 

the applied pressure comes from the expansion of a  

vapour  enclosed  in  either  a  rigid  stem  or  semi

-

rigid 

thermal system.

 

 

CERTIFICATIONS

 

 

Flameproof  models 

carry  the  following  label 

markings:

 

 

 

 

This  equipment  may  be  used  in  zones  1  &  2  with 

gases  groups  IIA,  IIB  &  IIC  and  dusts  groups  IIIA, 

IIIB & IIIC for temperatures classes T1, T2, T3, T4, 

T5 & T6.  

 

 

Intrinsic Safety Model

 

Input limitations for intrinsic safety:

 

Ui = 30V, Ii = 300mA, Pi = 0.6W

 

 

 

 

 

 

This equipment may be used in zones 0, 1 & 2 with 

gases  groups  IIA,  IIB  &  IIC  and  dusts  groups  IIIA, 

IIIB & IIIC for temperatures classes T1, T2, T3, T4, 

T5 & T6.  

 

 

Special Conditions of Use

 

Intrinsically safe models

 

1.

 

The  Sentry  Pressure,  Differential  Pressure 
and  Temperature  Switches  must  not  be 
installed in a location where moving dust flow 
can  generate  an  electrostatic  charge  on  the 
equipment. 

 

2.  The  Sentry  range  of  pressure  switches  may  be 
provided  with  an  aluminium  alloy  or  stainless  steel 
enclosure.  For  models  that  have  an  aluminium 
enclosure,  identified  in  the  model  number  by 
5xxxxxxxxxxxxxxxx  where  'X'  describes  other 
parameters of the equipment, the enclosure must be 
installed  in  such  a  manner  that  it  is  protected  from 
sources  of  impact  and  friction  when  installed  in  its 
end

-

use application.

 

INSTALLATION

 

 

Mounting (All models)

 

The  instruments  are  designed  to  be  mounted 

vertically  with  the  process  connection  underneath. 

However, mounting up to 45° from the vertical in any 

plane  is  acceptable,  although  a  small  calibration 

shift may occur. 

 