Introduction
UMSITRPDS3-1
November
2010
1-2
Appendix A, Online Configuration Map, is used to help navigate HART/transmitter menus during remote
configuration and operation of the transmitter.
Appendix B, Hazardous Area Installations, contains control drawings and other information needed for
installation in a hazardous area.
Appendix C, Elevation and Suppression Corrections, explains how to perform elevation and suppression
calculations necessary for certain liquid level gauging applications.
Warranty contains the product warranty statements and information concerning servicing of the product
during the warranty period.
1.2 PRODUCT DESCRIPTION
SITRANS P Series DSIII transmitters provide reliable, accurate, stable, and cost-effective measurement
of differential, absolute, and gauge pressure and liquid level.
The transmitter is a microcontroller-based, self-contained pressure-to-current transducer. A measuring
cell senses the applied process pressure and provides an analog output signal that is proportional to
applied pressure. An analog-to-digital converter produces a digital signal for the microcontroller. The
microcontroller modifies and corrects the signal for linearity and temperature, and a digital-to-analog
converter produces a 4-20 milliampere output signal for the loop.
The transmitter can be installed quickly and easily using one of the optional mounting brackets.
Illustrations showing mounting hardware and dimensions are provided in this manual. Measuring cell
construction determines a transmitter’s physical dimensions and mechanical installation. Note that a given
sensor construction can involve one or more measurement methods, as shown in Table 1-1.
TABLE 1-1 Measurements, Models and Figure References
Measurement
Model And Measuring Cell Construction
Refer To
Absolute or Gauge Pressure
7MF4033, Gauge Construction
7MF4233, Gauge Construction
7MF4333, Differential Construction
Figures 1-1, 4-3, 4-4, 4-
11, 4-13, 9-1, 9-2 and 9-
3
Differential Pressure
7MF4433, Differential Construction
7MF4533, Differential Construction
Figures 1-2, 1-3, 4-1, 4-
2, 4-12, 4-13, 9-3, and 9-
4
Liquid Level (Flange)
7MF4633, Flange Construction
7MF4812, Flange
Figures 1-4, 4-5, and 9-5
Figures 1-1 through 1-4 show the various transmitter models and constructions. They also show common
transmitter mounting methods. The mounting brackets shown are optional.
Содержание 7MF4033 series
Страница 48: ...Model 275 Universal Hart Communicator UMSITRPDS3 1 November 2010 2 26 ...
Страница 86: ...Installation UMSITRPDS3 1 November 2010 4 34 ...
Страница 152: ...Calibration and Maintenance UMSITRPSD3 1 November 2010 7 20 ...
Страница 158: ...Circuit Description UMSITRPDS3 1 November 2010 8 6 ...
Страница 197: ......
Страница 198: ......
Страница 200: ...Model Designations and Specifications UMSITRPDS3 1 November 2010 9 42 ...
Страница 204: ...Glossary UMSITRPDS3 1 November 2010 10 4 ...
Страница 208: ...Appendix A Online Configuration Map UMSITRPDS3 1 11 4 November 2010 ...
Страница 210: ...Appendix B Hazardous Area Installation UMSITRPDS3 1 12 2 November 2010 FIGURE 12 1 Control Drawing ...
Страница 211: ...UMSITRPDS3 1 Appendix B Hazardous Area Installation November 2010 12 3 FIGURE 12 2 Control Drawing ...
Страница 212: ...Appendix B Hazardous Area Installation UMSITRPDS3 1 12 4 November 2010 FIGURE 12 3 Control Drawing ...
Страница 216: ...Appendix C Elevation and Suppression Correction UMSITRPDS3 1 13 4 November 2010 ...