UMSITRPDS3-1
Model 275 Universal Hart Communicator
November 2010
2-13
2.5.1.1 New Configuration
This option is used to compile a custom set of device configuration data for downloading later to one or
more HART-compatible devices. Downloading the same data to multiple devices ensures that they all
store identical configuration data.
Use the following steps to compile off-line, new device configuration data:
1.
From the Main menu, press “1” to access the Offline menu.
2.
Press “1” to enter a new configuration. The Manufacturer
menu appears.
3.
Choose a manufacturer by scrolling to the manufacturer
name with the DOWN arrow, then pressing RIGHT
ARROW/SELECT. For a SITRANS P transmitter, choose
Siemens. The Model menu appears.
4.
From the Model menu, choose a device by scrolling through
the list, then pressing RIGHT ARROW/SELECT. The Field
Device Revision (Fld dev rev) menu appears.
The Field Device Revision menu contains the currently
installed software revisions for the field device and device
descriptions (DD) for the model selected from the Model
menu.
Select the software revision (RIGHT ARROW/SELECT or
number) to access the Blank Template menu (at right). To
display the software revision for a particular device, connect
the Communicator to the device and follow instructions
given in the device manual.
To find the field device revision (compatible device
description) for a SITRANS P transmitter:
1.
Connect the Communicator to the transmitter.
2.
Apply power to the transmitter.
3.
Turn on the Communicator.
4.
In the Online menu, select “3 Device Setup.”
5.
In the Device Setup menu, select “4 Detailed Setup.”
6.
In the Detailed setup menu, select “4 Device
Information.”
7.
In the Device Information menu, select “6 Revision
numbers.” Revision numbers for the Communicator and
the transmitter will be listed.
HART Communicator
Offline
"
1->New configuration
2 Saved configuration
HELP
Unnamed
From Blank Template
"
1->Mark all
2 Unmark all
3 Edit individually
4 Save as...
HELP |SAVE
Содержание 7MF4033 series
Страница 48: ...Model 275 Universal Hart Communicator UMSITRPDS3 1 November 2010 2 26 ...
Страница 86: ...Installation UMSITRPDS3 1 November 2010 4 34 ...
Страница 152: ...Calibration and Maintenance UMSITRPSD3 1 November 2010 7 20 ...
Страница 158: ...Circuit Description UMSITRPDS3 1 November 2010 8 6 ...
Страница 197: ......
Страница 198: ......
Страница 200: ...Model Designations and Specifications UMSITRPDS3 1 November 2010 9 42 ...
Страница 204: ...Glossary UMSITRPDS3 1 November 2010 10 4 ...
Страница 208: ...Appendix A Online Configuration Map UMSITRPDS3 1 11 4 November 2010 ...
Страница 210: ...Appendix B Hazardous Area Installation UMSITRPDS3 1 12 2 November 2010 FIGURE 12 1 Control Drawing ...
Страница 211: ...UMSITRPDS3 1 Appendix B Hazardous Area Installation November 2010 12 3 FIGURE 12 2 Control Drawing ...
Страница 212: ...Appendix B Hazardous Area Installation UMSITRPDS3 1 12 4 November 2010 FIGURE 12 3 Control Drawing ...
Страница 216: ...Appendix C Elevation and Suppression Correction UMSITRPDS3 1 13 4 November 2010 ...