Maint 06-0080
(2) About White Level calibration
Calculates a ratio between White Background part on the CIS Unit assembly and color level of the OCR TEST
SHEET scanned actually, and calibrates color image level inputted the scanner with the ratio.
Determination value: within the range of 898 - 1455
(3) About Feeding Speed calibration
Marks at 11 locations in the OCR TEST SHEET are scanned. The velocity error for the feeding direction is
learned based on the reference values and the measured values. The velocity adjustment coefficients are
learned based on that.
Determination value: withing ±1%