4. Gas Control Module (Enhanced)
70
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
2016-04
Switching OFF the Gas Control Module
At the end of the working day, when measurements are finished and the GCM
enhanced is no longer needed, switch off the GCM using the ON/OFF switch on
the rear panel.
Figure 35
Close the valves of the gas cylinder or laboratory gas handling system, remove
the tubes from the GCM and let out any remaining gas from the tubing.
Data-logging
The
i-control
software supports data-logging and data display for the
GCM
enhanced
.
For information about data-logging (e.g. prerequisites, driver installation,
connecting to GCM Enhanced, importing logged data, data displayed in status
bar, data displayed in Excel output or precautions before starting a measurement)
see IFU i-control software, chapter 6. Gas Control Module (GCM) Enhanced
Support.