Result Display in MS Excel
The MS Excel results sheet generated by the i-control software displays a single
average measurement value for each well that has been measured using the
Optimal Read function. The employed Optimal Read settings, i.e. the overall
number of flashes as well as the number of flashes per well, are also displayed.
Figure 39:
Results output for a measurement with Optimal Read (example for a
96-well plate)
Miscellaneous Features of Optimal Read
Optimal Read is only available for Fluorescence Intensity Bottom measurements
in combination with the Enhanced FI Bottom module (4 mm FI Bottom fiber).
The Optimal Read feature is not available when performing well-wise
measurements.
The standard MRW function for Fluorescence Intensity Bottom reads is disabled
when “Optimal Read” is activated and vice versa.