5. Operating the Instrument
120
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
2016-04
Injection Mode
The injection settings can be adjusted via the software:
Injection
Select Injector:
Injector A or Injector B can be selected. It is not possible to select both injectors
on one strip. If a measurement with two injectors is to be performed, two injector
strips are necessary.
Speed
: The injection speed is selectable from 100 – 300 µl/sec for each
injector.
Select a ‘
Refill speed
’ from 100 – 300 µl/sec. for each injector or check the
‘Refill Speed equal to Injection Speed’ box.
Select refill mode ‘
Standard
’ if refill should be performed when syringe is empty
(multiple injection steps are performed before refilling, refill occurs after
dispensing approx. 800 µl). Select ‘
Refill for every injection
’ if refill should be
performed for every injection step.
Injection volume
The injection volume depends on the microplate type. The plate definition files
include a so-called working volume. This working volume defines the maximum
volume to be injected into the selected microplate. Therefore, always make sure
that the selected plate definition file contains a correct setting for the working
volume. The maximum injection volume is 800 µl/injection strip. If volumes
greater than 800 µl are to be injected (e.g. into 6-well plates), more than one
injection strip has to be used.