Luminescence Optics
In luminescence measurement mode, the Infinite 200 PRO uses fixed microplate
position and a moveable luminescence measurement head (see
Optical System Luminescence). The plate thickness is defined by selecting the
corresponding plate type in the software (see i-control Instructions for Use).
Fiber
A glass fiber guides the light from the sample to the detection unit. The fiber is
designed to measure from 6-well up to 384-well plates.
Filter Wheel
A filter wheel with 6 filter positions in front of the PMT window is switched to the
required luminescence channel. The sensitivity of the detection system makes it
necessary to attenuate high luminescence light levels; therefore, the filter wheel
can also switch a neutral density filter across the selected fiber exit.
Filter Wheel Position
Filter
Position 1
Lumi Green*
Position 2
Lumi Magenta*
Position 3
OD2 neutral density filter
Position 4
No attenuation
Position 5
Blue 1**
Position 6
Green 1**
*
recommended for the BRET
2
assay and the ChromaGlo - Luciferase Assay
** recommended for the BRET assay
The OD2 neutral density filter serves to attenuate high light levels by a factor of
100 (corresponding to 2 OD absorbance). The resulting values are automatically
scaled to counts per second and displayed accordingly in the software results
output.
See Figure 16 to Figure 19 for transmission spectra of luminescence filters.