5. Operating the Instrument
118
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
2016-04
5.11.3 Before Starting a Measurement with Injectors
Before starting a measurement make sure that:
•
The tubes are clean. If not please refer to chapter 5.10.2 Priming and
Washing of the Infinite 200 PRO for details how to clean the injector system.
•
The injector tubes are correctly inserted into the storage bottles and fixed.
•
The injector system is primed. It is not possible to start a measurement
without priming the system.
When priming the system:
•
Check the tubes for leaks, visually or using a nonhazardous fluid, before
priming with valuable reagents.
•
Check the tubes for kinks, visually or using a nonhazardous fluid, before
priming with valuable reagents.
•
Make sure that the injector needles are not twisted.
•
If the tubes require replacement for any reason, after the tubes have been
changed do not forget to perform washing and priming before starting a
measurement.
5.11.4 Injector Modes (i-control)
When using the injector, two modes are available:
Dispense
: The dispense mode allows liquid to be dispensed plate-wise into the
selected wells
Injection
: This mode must be used in combination with a measurement strip. The
injection is performed in a well-wise mode.
Dispense Mode
The dispense settings can be adjusted via the software:
Dispense
Select Injector
: Injector A and/or Injector B can be selected.
Speed
: The injection speed is selectable from 100 – 300 µl/sec for each
injector.
Select ‘
Refill speed
’ from 100 – 300 µl/sec. for each injector or select
‘
Refill Speed equal to Dispense Speed
’.
Select refill mode ‘
Standard
’, if refill should be performed when syringe is
empty (multiple dispense steps are performed before refilling, refill occurs after
dispensing approx. 800 µl).
Select ‘
Refill for every dispense
’ if refill should be performed for every
dispense step.