7. Quality Control
2016-04
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
175
Measurement Parameters:
Parameter
Setting
Reading Mode
AlphaScreen/AlphaLISA
Ex Wavelength
680 (30) nm
Em Wavelength
570 (100) nm
Excitation Time
1000 ms
Integration Time
500 ms
Gain
Calculated from Well: A1
Plate Type
GRE384fw
Evaluation:
Calculate the average and standard deviation for each Omnibead concentration.
Perform a blank reduction by subtracting the average signal of the blanks wells
from the average signal of each Omnibead concentration.
Plot the average blank-corrected values against the final Omnibead
concentrations in a XY scatter diagram. Add a linear trend line with intercept set
to 0 and solve the trend line equation (
y = kx
) using the 3-fold standard deviation
of the blank as y.
k
y
x
=
y
3* stdev of the blank (wells N2-P6)
Extrapolate the detection limit [ng/ml] by using the 3-fold standard deviation of the
blank as y.