4. Gas Control Module (Enhanced)
62
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
2016-04
4.5
Installing the Gas Control Module
The gas control module is intended for operation with the Infinite 200 PRO reader
under controlled laboratory conditions. Before installing and switching on the gas
control module, ensure that the designated site meets the following requirements.
4.5.1
Requirements
Environmental Requirements
•
Operate the gas control module in a well-ventilated, temperature and
humidity controlled (air-conditioned) environment.
Temperature:
15 °C (59 °F) – 30 °C (86 °F)
Humidity:
20 – 80 % non-condensing
•
Do not expose or locate the gas control module near direct sunlight or heat
sources.
•
Maintain a low-dust environment. Keep liquids and vapors away from the
gas control module.
•
Adequate ventilation must be provided for the room in which CO
2
and N
2
are
used.
Space Requirements
•
Place the gas control module onto the Infinite 200 PRO or on a rigid, level,
and vibration-free surface near the instrument. The GCM is 18.0 cm (7.1 in)
high, 25.0 cm (9.8 in) wide, 17.0 cm (6.7 in) deep and weighs 1.5 kg
(3.3 lbs).
•
Leave sufficient distance behind the gas control module for access to the
rear panel. Make sure that all gas tubes, the power cable, and data cable
are accessible and in no way obstructed.
Power Requirements
•
The power supply of the gas control module is auto-sensing and able to
operate without any manual adjustments within 100 - 240V (AC), 50/60Hz.