7. Quality Control
172
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
2016-04
7.3.9
Absorbance Baseline Flatness (Infinite F200 PRO)
Perform the following measurements to determine the baseline-flatness with the
available filter:
Measurement Parameters:
Parameter
Setting
Reading Mode
Absorbance
Measurement Wavelength
340 nm
405 nm
492 nm
590 nm
620 nm
700 nm
Number of flashes
25
Settle Time
0
Plate Type
GRE96ft
Part of the Plate
A1
Kinetic Cycles
20, Minimal Interval Time
Plate Layout:
No plate is necessary for measurement – the plate carrier has to be empty for this
measurement.
Material/Reagents:
No material or reagents necessary for this test.
Calculation of Baseline Flatness:
Calculate the standard deviation over 20 cycles for each wavelength.