5. Operating the Instrument
128
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
2016-04
Example 3:
Usage of ‘Move Cuvette OUT’ strip when measuring a combination of microplate
and cuvette:
Cuvette
Necessary for cuvette measurement
Absorbance strip
(cuvette)
Up to 4 absorbance fixed wavelength strips are allowed. Reference
wavelength is only selectable when using one absorbance fixed
wavelength strip. ‘Apply blanking’ is disabled when reference
wavelength is selected.
Select the appropriate measurement parameters (wavelength,
number of flashes and settle time)
Move Cuvette (Out)
The cuvette holder is moved ‘out’.
User Request
The user request interrupts the measurement and therefore allows
removing the cuvette from the cuvette port. When confirming the
request the measurement continues.
Move Cuvette (In)
The cuvette port is moved in.
Plate
Necessary for microplate measurements. Select an appropriate plate
type for the measurement.
Part of Plate
(not shown)
Optional. Use the ‘part of plate’ strip if only a part of the plate will be
measured.
Absorbance strip
(microplate)
Up to 10 absorbance fixed wavelength strips are allowed. Reference
wavelength is only allowed on the first absorbance strip. Reference
wavelength is disabled on absorbance strips 2 to 10.
Select the appropriate measurement parameters (wavelength,
number of flashes and settle time) for your application.
Move Plate
Optional. To move the microplate automatically out of the instrument
when finishing the measurement, select the ‘Move plate OUT’.