7. Quality Control
2016-04
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
157
7.3
Specifications - Test Instructions
7.3.1
Fluorescence Top
For the Infinite 200 PRO with the option ‘Fluorescence Top’, the following tests
can be performed to prove the specifications:
•
Sensitivity
•
Uniformity
•
Precision
These test instructions are valid for the Infinite 200 PRO:
•
Infinite F200 PRO
•
Infinite M200 PRO
•
Standard version
•
Spectrally enhanced version
Sensitivity
Perform the following measurement to determine the detection limit for
Fluorescein:
Measurement Parameters:
Parameter
Setting
Reading Mode
Fluorescence Top
Ex Wavelength
Infinite F200 PRO: 485 (20) nm
Infinite M200 PRO: 485 nm
Em Wavelength
Infinite F200 PRO: 535 (25) nm
Infinite M200 PRO: 535 nm
Number of flashes
25
Integration Time
40
Settle Time
0
Gain
Optimal
Plate Type
GRE96fb