9. Troubleshooting
2016-04
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
187
Error #
Error Text
Description
39
Injector A Communication
Timeout
Communication Error on Injector Interface. Please report
this error to your local customer support office.
40
Injector B Communication
Timeout
Communication Error on Injector Interface. Please report
this error to your local customer support office.
41
Prime Wash Error
Injectors still priming or washing. Please wait until prime or
wash process is finished.
42
Instrument is locked
Instrument is locked after a serious hardware problem. For
unlocking a reboot is necessary. Please report this error to
your local customer support office.
43
Prepare: [channel]:
Wavelength:[lambda]
Gain:[g], Counts: [cts]
Unspecific Hardware failure. Please report this error to your
local Tecan customer support office.
44
Steploss Error
Actuator failure. Please report this error to your local Tecan
customer support office.
45
Sync Scan: Number of EX-
Steps does not match EM-
Steps
Unspecific error in the Instrument - Computer
communication protocol. Please report this error to your
local Tecan customer support office.
46
Handshake timeout at
module
Unspecific Hardware Failure. Please report this error to
your local Tecan customer support office.
47
Motor Timeout
Unspecific Hardware Failure. Please report this error to
your local Tecan customer support office.
48
[Value] is not in defined a
Range
Unspecific Hardware Failure. Please report this error to
your local Tecan customer support office.
49
Sensor is broken
Sensor Failure. Please report this error to your local Tecan
customer support office.