7. Quality Control
2016-04
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
167
7.3.4
Fluorescence Polarization (Infinite F200 PRO only)
For the (Infinite F200 PRO)
with the option ‘Fluorescence Polarization’ the
following tests may be performed to prove the specifications:
•
Precision/Repeatability
These test instructions are valid for the (Infinite F200 PRO):
•
Standard version
•
Spectrally enhanced version
Precision
Perform the following measurement to determine the precision:
Measurement Parameters:
Parameter
Setting
Reading Mode
Fluorescence Polarization
Ex Wavelength
485 (20) nm
Em Wavelength
535 (25) nm
Number of flashes
25
Integration Time
40
Lag Time
0
Time between Move and Read
0
Gain
Optimal
Plate Type
GRE96fb
Reference from/to
A1 – D1
Reference blank from/to
A2 – D2
Reference value
20 mP
Measurement blank from/to
same as reference blank