7. Quality Control
162
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
2016-04
Material/Reagents:
25 nM Fluorescein (in 0.01 M NaOH) (Fluorescein sodium salt, Sigma F6377)
0.01 M NaOH (=Blank) (NaOH pellets, Merck article no. 6495 or Sigma S8045)
1 Greiner 96-well plate µClear, black with transparent bottom
200 µl P tips
Calculation of Detection Limit (Sensitivity):
See 7.3.1 Fluorescence Top: Sensitivity.
Sensitivity (Enhanced)
Perform the following measurement to determine the detection limit for
Fluorescein:
Measurement Parameters:
Parameter
Setting
Reading Mode
Fluorescence Bottom
Ex Wavelength
Infinite F200 PRO: 485 (20) nm
Infinite M200 PRO: 485 nm
Em Wavelength
Infinite F200 PRO: 535 (25) nm
Infinite M200 PRO: 535 nm
Number of flashes
25
Integration Time
40
Settle Time
0
Gain
Optimal
Plate Type
GRE96fb
Plate Layout
Pipette 200 µl of 1 nM fluorescein or the blank solution (0.01 M NaOH) into the