5. Operating the Instrument
2016-04
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
75
5. Operating the Instrument
5.1
Introduction
The Infinite 200 PRO is operated using a personal computer based software
control.
i-control
or
Magellan
software may be used as the user interface. For
details see the corresponding software Instructions for Use. This short
introduction is for a general understanding of instrument parameters and
operation. Suggestions are made on how to optimize instrument parameters for
your applications.
Every effort has been made to ensure that the instrument will work correctly even
if the default parameters are not appropriate for a particular application - with an
important exception:
CAUTION
WHEN PLACING A MICROPLATE INTO THE PLATE CARRIER, ALWAYS
MAKE SURE THAT THE CORRECT PLATE DEFINITION FILE (PLATE
HEIGHT) HAS BEEN SELECTED IN THE SOFTWARE BEFORE YOU DO
ANYTHING ELSE.
MAXIMUM PLATE HEIGHT IS 23 MM (INCLUDING LID).
CAUTION
BEFORE STARTING MEASUREMENTS, MAKE SURE THAT THE
MICROPLATE POSITION A1 IS INSERTED CORRECTLY. THE POSITION
OF WELL A1 HAS TO BE ON THE UPPER LEFT SIDE.
CAUTION
IN CASE OF SIGNIFICANT SOILING OF THE PLATE TRANSPORT, THE
SPRING MECHANISM MIGHT NOT WORK PROPERLY, AND CAN LEAD
TO WRONG POSITIONING. PLEASE CONTACT YOUR LOCAL SERVICE
CENTER.
IMPORTANT
WHEN OPERATING THE
INFINITE 200 PRO
ALWAYS WORK
ACCORDING TO GLP GUIDELINES.
CAUTION
THE
INFINITE 200 PRO
HAS A FAN ON THE BACKSIDE OF THE
INSTRUMENT THAT DRAWS IN AIR. THE AIR FILTER HAS TO BE
CHECKED EVERY 4 WEEKS AND BE REPLACED WHEN DIRTY. THE
AIR FILTER MUST BE REPLACED AFTER 6 MONTHS.
STOP
STOP
STOP
STOP