7. Quality Control
166
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
2016-04
Material/Reagents:
1 nM Europium (B119-100, HVD Live Sciences)
Enhancement Solution (=Blank) (1244-105, HVD Live Sciences)
1 Greiner 96-well plate white
200 µl P tips
Calculation of Detection Limit (Sensitivity):
See 7.3.1 Fluorescence Top: Sensitivity.
Precision
Perform the following measurement to determine the precision/reproducibility:
Measurement Parameters:
Parameter
Setting
Reading Mode
Fluorescence Top
Ex Wavelength
Infinite F200 PRO: 340 (35) nm
Infinite M200 PRO: 340 nm
Em Wavelength
Infinite F200 PRO: 612 (10) nm
Infinite M200 PRO: 617 nm
Number of flashes
25
Integration Time
400
Lag Time
100
Time between Move and Flash
0
Gain
Optimal
Plate Type
GRE96fw
Part of the plate
A1
Kinetic
20 Cycles
Interval Time
Minimal
Plate Layout:
See 7.3.3 Time Resolved Fluorescence Precision.
Calculation of Precision: